| DIMENSIONS |
Logger
140mm x 126mm x 17mm |
Thermal Barriers
TB20 - 20mm X 175mm X 175mm
TB24 - 24mm X 175mm X 175mm
TB40 - 40mm X 180mm X 220mm
TB60 - 60mm x 210mm x 250mm
TB80 - 80mm x 220mm x 260mm
TB120 - 120mm x 260mm x 300mm
TBSPECIA - special version with water coil |
| THERMACOUPLE INPUTS |
Range
Tipo E: -270°C +1000°C
Tipo J: -270°C +1000°C
Tipo K: -270°C +1372°C
Tipo T: -250°C +400°C
Tipo N: -210°C +1200°C |
Resolution
0.1°C |
Accuracy
±0.5°C (included compensation of cold junction) |
| VOLTAGE INPUTS |
Range ±19.5mV ±2.5
V in 8 step |
Resolution
600 nV - 76uV in 8 step |
Accuracy
0.01% FS or 50uV, whichever the greater |
| CURRENT INPUTS |
Range +195uA to +25mA in 8 steps (100ohm sense resistor) |
Resolution
6 nA to 763 nA in 8 steps (100 ohm sense resistor) |
Accuracy
0.01% or 500 nA, whichever is greater |
| NOISE REDUCTION |
Common Mode
100dB (sampling interval 2s or more)
Common Mode 50/60 Hz
150dB (sampling interval 2s or more)
Normal Mode 50/60 Hz
100dB (sampling interval 2s or more) |
| BATTERY |
Lithium Thionyl Chloride, 3.6V
Expected life equivalent to one test per day for five years. |